Defibrillator Specifications
General
- Display : 320 x 240 pixels, 5.7” Blue LCD with LED backlit
- Display area : 120mm x 90mm
- Dimensions : H180 x W230 x D210mm (max.) without Paddle
- Weight : 4.5 Kgs.
Defibrillator
- Operating Mode : Manual
- Waveform : Biphasic
- Energy : 2 to 200 joules in 12 steps
- Synchronous : Energy delivery begins within 60 ms of the QRS Peak
- Charging Time : Less than 8 seconds
- Paddle Assembly : External paddle assembly with coil retractable high voltage cable
- Adults & Pediatric paddle integrated to same handle
- Check Mode : Manual test mode
- Operation : By front Panel keyboard and paddle buttons
ECG:
- Frequency Response : 0.5 Hz to 40 Hz (ECG)
- Frequency Response : 0.5 Hz to 30 Hz (Paddle)
- Patient isolation : Optical isolation
- Lead selection : Lead I,II and III
- Gain setting : 0.5, 1.0 and 2.0
- CMRR : > 90 db
- Sweep Speed : 12.5, 25.0, 50.0 mm/sec
- Waveform 1 : ECG
- Waveform 2 : 4sec delayed (cascaded)
- Data Output : Analog output of ECG
- Digital filter : Digital signal processing for good quality of ECG
- Freeze mode : To freeze last 4 seconds ECG
Heart Rate
- Range : 30-250 BPM
- Accuracy : +/- 2 BPM
- Resolution : 1 BPM
- Tone : Beep on every QRS detected
- Low alarm Limit Range : 30-245 (step of 5)
- High alarm Limit range : 40-250 (step of 5)
Printer
- Paper Type : Thermal paper 50/55 mm width, 20 meters length
Event Printing
- ECG Mode : Prints 5 Seconds pre & post ECG
ECG in Defib
- Mode : Prints 5 Seconds pre & post ECG
- Defib Mode : Stores & prints 10 Seconds of Defib Shock data i.e. 5 seconds pre critical & 5 seconds post critical Maximum events that can be store Memory of 10 Manual Defib
- Check Mode : Prints the status of Check Mode
- Print Annotates : Time, Date, ECG Lead, Mode, Shock Status & selected Energy
- Delay : 10 Seconds
Power Supply (AC Mains Through Adaptor)
- Input : 150-276VAC, 50Hz, 60Hz, 12V DC for ambulance
- Output : 17V@4A
- Battery : 12V/3.8 Ah Rechargeable
- Back-up time : 3 hours min on full charge in ECG mode /50 shock of 150 Joules in defib mode
- Power consumption : 30 Watts (max) in ECG mode
Syringe Pump Features
· User Friendly Interface on 2.6 inches HD LCD Display
· Accepts 5, 10, 20, 30, 50, 100 mL syringe
· Mode of operation: Simple, Vol/Time, Vol/Weight, RC
· RC mode allows user to change rate during ongoing infusion
· 0.1mL/Hr to 1500 mL/Hr flow rate
· Programmable to accept any brand of syringe
· Target volume and Time setting
· Elapsed time display
· KVO (Keep Vein Open) automatically opens as infusion is complete
· Front Loading Spill Proof System
· Freely Stackable
· Automatic syringe size detection
· Variable Bolus/ Purge function
· Infused volume display
· Variable occlusion pressure
· Built in rechargeable battery
· Connection to Nurse Call & 12V supply (optional) for Mobile/ Ambulance Applications
· Anti-siphon protection
· Anti Bolus System (ABS) (Optional)