Application |
Instruments and Devices that are sensitive to the high
temperatures of steam sterilization are sterilized by Low Temperature 100%
Ethylene Oxide Gas. |
Chamber
Sizes / Volume / EO Cartridges. |
· “Chamber
Size A” 430 (W) x 380 (H) x 0830 (D). Volume
140 Liters (5 cubic feet) &100gms. ETO Cartridge. · “Chamber
Size B” 510 (W) x 460 (H) x 0970 (D). Volume
225 Liters (8 cubic feet)&170gms. ETO Cartridge. · “Chamber
Size C” 600 (W) x 600 (H) x 1200 (D). Volume 430 Liters (16 cubic feet) &300gms.
ETO Cartridge. |
ETO
as sterilant |
· Ethylene
Oxide is highly penetrating sterilant favorable for complex instruments and
is compatible with most device materials. · The
ETO sterilization is safe, effective and economical process for Low
Temperature Sterilization. · ETO
gas is gentle on device materials as compared to Hydrogen Peroxide (Oxidative
Sterilant). Thus sterilization by ETO gas extends the life of devices and
needs fewer repairs. · 100%
ETO gas sterilization is the most cost effective method as compared to
vaporized Hydrogen Peroxide system. |
ETO
cartridge |
Single
unit-does EO cartridge is punctured inside the negative pressure chamber only
when proper chamber conditions are achieved. |
Temperature
Monitoring |
Continuous
temperature monitoring maintains temperature within +/-3⁰C of the selected
cycle temperature and automatically shuts off if the temperature reaches 4⁰C
beyond set point. |
Flexibility,
Access code & Reliability. |
Optimal
flexibility in programming the recipe to suit a wide variety of products
& Programmable access codes allow you to monitor and trace cycle
operator.Accuracy, reliability and repeatability of process parameters. |
Lumen
Restrictions |
No
lumen bore restriction’s |
Chamber
Fabrication. |
The rectangular chamber
is designed for full vacuum. This takes care of requirements of EO
sterilization process.The rectangular chamber has round corners which
facilitates cleaning as specified by cGMP. Rectangular concave gaskets
suitable for chamber give more consistent results over prolonged period minimizing
leakage under vacuum. |
Single
Hinged Door. |
The chamber has a single
hinged door on one side to facilitate easy loading and unloading. Welded
dished end cover is provided at the other end. |
Insulation |
The insulation provided
for chamber, CFC free “Polyisocyanurate” of adequate thickness. |
Ventury
VacuumSystem |
A venturyvacuum system
is provided along with the unit for proper air removal in the early stages of
the cycle as well as removal of residual gas at the end of the cycle. |
PLC
Control System. |
·
PLC control
system. · Powerful
and versatile PLC for process control. · Coloured
7 inch HMI with necessary displays. · Control
panel with electrical switchgear. ·
Manual back up. |
Recorder. |
A compact 40 column thermal printer is
provided for batch record of temperature/ pressure & time. |
Sterilization
Cycles |
· Warm
Cycle at 55⁰C. Cycle time 4 hours. · Cool
Cycle at 37⁰C. Cycle time 6 hours. · Aeration
Time for 55 0 C cycle is 12 hours and for 370C cycle is
24 hours. |
Safeties |
· The
process will not start unless the door is properly closed. · If
chamber absolute pressure increases beyond set level, cycle will not
continue. · Emergency
off when required. · Maximum
operator safety. |
Electricity |
220 Volts, 50 Hz,15 Amps circuit |