Syringe Size: 2ml, 5ml, 10ml, 20ml, 30ml, 50/60ml
Volume Infused: 0.1~9999.9 ml, Step by 0.1 ml
Accuracy: ±2%
Flow Rate
2ml:0.1-100ml/h; 5ml:0.1-150ml/h;
10ml:01:400ml/h; 20ml:01-600ml/h;
30ml:0.1: 1000ml/h; 50/60ml:0.1-800ml/h;
Bolus Rate
2 ml Syringe: 100 ml/h; 5 ml Syringe: 150 ml/h;
10 ml Syringe: 400 ml/h; 20 ml Syringe: 600 ml/h;
30 ml Syringe: 800 ml/h; 50/60 ml Syringe: 800 ml/h
KVO Rate
Flow rate ≥ KVO rate 3 ml/h
Flow rate ≥ 1ml/h and <10ml/h: KVO rate 1 ml/h
Flow rate < 1ml/h: KVO rate = Flow rate
Alarm
Battery exhausted alarm, Low battery alarm, Near end alarm,
VTBI finished alarm, Incorrect syringe handle installation alarm,
Syringe size error, Plunger error, Pause overtime alarm,
Occlusion alarm, Malfunction alarm, A.C. and D.C. Off
Occlusion Pressure
Two levels: High, Low
H: 100 kpa; L: 50 kpa
Power: A.C. 100-240V 50/60Hz, Input D.C. 15V/2A Output, Input Voltage to Device
Power and Battery: 55VA Max; Battery Voltage 10.8V; Support 6 Hours normal operation
Classification: Class II, Type CF, IPX2
Dimension: 278x145x106mm (WxDxH); 1.5kg