Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
· Easy to use interface ,12.1” Color TFT display with Touch Screen Key pad/Touch Screen Lock facility for protection against accidental/un intentional changes
· Simultaneously display of Three Waveforms & Three Loops
· Multi level mechanism for patient safety with intelligent high alert alarming system, hence increase safety of operation Three Level color coded audio-visual alarms
· Maintenance - free, auto self testing and no need routine calibration of flow sensor
· Fixed reusable flow sensor hence no daily recurring cost
· Lowest in class bias flow 2.0 L/min, Low o2 consumption
· Automatic Tube & leakage compensation technology
· Advance Pre Test System before use , assured the professions to safely use of machine on patients
· Medical air compressor (Model AF-12) option ensures a constant, compressed air supply in case main central line fail. Unique anti vibration design assures low noise & quiet operation with sound level
· Air Compressor driven technology which makes, more accurate in spontaneous delivery
· Humidifier option ensure a high level of gas humidification to patient
· Anti condensation heated exhalation system with auto sterilization can (Make ICU infection free)
· Auto-Cleavable at 134 degree can use safely and prevent cross infection
· Low maintenance cost, Easily removable, trouble free Active Exhalation Valve technology.
· Up gradable to Capnography (EtCO2) &Pulse Oximetry (Spo2)*
Adult-Pediatric Ventilation
- Tidal volume from 10ml to 2000ml (Pediatric-Adult) Pressure control mode with tidal volume security/guarantee provide the advantage of volume control ventilation both invasive & non invasive ventilation
Neonatal Ventilation
- Continuous Flow based (VIVE) Proven Technology for neonatal patients helps flushing dead space in Y piece and maintaining PEEP
- Proximal digital Flow Sensor option provides precise measurement of volume & flow directly at airway, a quick response time & less work breathing . It provides tidal volumes as low as 2 ml with the Neonatal option for effective, Safe ventilation even for the smallest patients.
Powerful monitoring functions
- Lung mechanics parameters
- Up to 1500 Alarm Events review
- 72-hours Numerical & Graphical trend review
- Saved loops with comparison review*
- 10-Hours Breath to Breath Waveform review*
Utilities
- Inspiratory Hold & Expiratory Hold
- 100% O2 flushing
- Manual Breath
- SIGH Breath (Configurable)
- Nebulizer- Nebulization with tidal volume compensation (1-30 min)
Tools (Optional)*
- PV Tools (measure LIP & UIP)*
- MIP/NIF (Maximum Inspiratory Pressure)*
- PO.1 (Occlusion Pressure)*
SmartVent (Automatic Ventilation Mode)
Intelligent mode of ventilation-Easy to Use & lung protective
Ensure the guaranteed minute ventilation breath by breath at lowest inspiration pressure and helps patient for spontaneously activity from start of ventilation
Automatically set Parameters according to Body Weight
Auto Synchronous the flow on patient demand, Improve Patient Outcome, and avoid ventilator fighting on Spontaneous
Results shorter ventilation times
PRVC ( Pressure Regulated Volume Control )
Advanced Dual Modes Make it more Comfortable to patient Wean Case to Case
Deliver Preset Tidal Volume at Same Rate with the Lowest Possible Pressure
DuoVent/ APRV Mode (Resistance Free Ventilation)
Provide lung protective ventilation, Use an “open lung” approach including alveolar recruitment, improved oxygenation. in ARDS patients.
Pressure support setting in both lunge for spontaneous breathing to patients.
NIV S/T ( Non-Invasive with Synchrony Feature )
More Unique Benefits , High Performance Non Invasive Ventilation. Also Set Trigger, Rate with Additional Parameters
nCPAP/nBPAP
Most effective therapy , used worldwide , It also recommend to use NIV option in newborns.
VG Option ( Volume Guarantee)*
New Approaches in volume control ventilation for neonatal have Potential advantages of less risk for barotraumas becouse the tidal volume delivered is limited Reduced peak pressure especially when the baby makes a significant contribution from spontaneous effort
Adult |
Pediatric |
Neonatal (Optional) |
|
Modes |
VCV, SIMV-VC, PCV,SIMV-PC, PSV, CPAP+Apnea Backup, APRV, DuoVent, PRVC, SIMV-PRVC, SmartVent (Optional) Non-Invasive: NIV, NIV-ST, O2 Therapy |
VCV, SIMV-VC, PCV, SIMV-PC, PSV, CPAP+Apnea Backup, APRV, DuoVent, PRVC, SIMV-PRVC, SmartVent (Optional) Non-Invasive: NIV, NIV-ST, O2 Therapy |
IPPV, SIPPV, SIMV, PSV, CPAP, VCV Non-Invasive: nCPAP, nBPAP O2 Therapy |
Tidal Volume |
150-2000 ml |
10-350 ml |
2-100 ml |
Frequency |
1-100 bpm |
4-150 bpm |
4-200 bpm |
Apnea Back Up Frequency |
4-50 bpm |
4-80 bpm |
2-100 bpm |
Apnea Time |
3-45 Seconds |
3-45 Seconds |
3-45 Seconds |
O2 Therapy Flow |
1-60 L/min |
1-50 L/min |
1-40 L/min |
Inspiration time |
0.1- 12 S (increment:0.1S) |
0.1- 12 S (increment:0.1S) |
0.1- 3 S (increment: 0.1S) |
T_Pause/T_Rise |
0-50% /Auto |
||
FiO2 |
21%-100% , 100% O2 (0-5 minutes) |
||
Trigger |
Flow Trigger :0 to 20 L/min, Pressure Trigger : – 20 to 0 cm H2O,ETS: % 5 to 80% |
||
Flow range/Pattern |
0-120 L/min, Square, Decelerating |
||
Pressure Control |
5-120 cmH2O |
||
Pressure Support |
0-120 cmH2O |
||
Electronic PEEP |
0-50 cm H2O (increment:1 cm H2O) |
||
I:E Ratio |
1:9.9 – 9.9:1 |
||
Nebulizer |
0-30 minutes (Particle Size <_ 3 micron) Optional |
||
Waveforms & Loops |
Waveforms : P-T, F-T, V-T , Etco2* , Spo2* & Loops : P-V, F-V, P-F |
||
Recall / Review |
1500 alarm events , 72 Hours Numerical & Graphical Trends, Saved loops with comparison (optional)*, 10 Hours Breath to Breath Waveform Review (Optional)* |
||
Monitoring Parameters |
Ppeak, Pmean, Pplat, PEEP, VTi, VTe, MVe, MV_cont., MV_spont, F_Peak_ insp, F_ Peak_ Exp, Leak%, Freq, F_cont., F_spont, % F spont, I:E, Ti,Te,Ti/T total, FiO2, (Spo2,Etco2-optional)* Cstat, Cdyn, Rins, Rexp, RSBI, WOB, Auto PEEP, MIP, P0.1, TC, C/C20 |
||
Alarm |
Three Level Audio-visual color coded alarms includes VTe high/low, MV high/low, Tidal volume not reached, High leak Peak Pressure high/low, Pressure limitation, O2 supply pressure high/low/fail, Air supply pressure high/low/fail, Apnea, Apnea Backup, Freq high/low, FiO2 high/low, Circle Block, Circuit disconnection, Expiatory block Power: AC Power Fail, Low Battery, Battery too Low |
||
GENERAL |
|||
Dimensions/Weight |
285x350x390 mm (HxWxD) / 12Kg |
||
Communication Interface |
VGA, USB, RJ45 (optional), CO2,SpO2. |
||
Display |
12.1 inch color TFT with Touch Screen (Ratable) |
||
Air/Oxygen Gas supply |
2.4 to 6.0 bar central supply or integrated Air Compressor (AF-12)-Optional-Humidifier |
||
Power Supply |
100-265V AC, 50/60Hz, 1A, 100 VA ,Battery 9.6V,9Ah Lithium Ion Phosphate ,(LiFePO4) Backup > 180min. |
||
SAFETY STANDARD |
|||
Basic Construction |
In accordance with IEC 60601-1, IEC 60601-1-2 & EMC 60601-2 |
||
Unit Type |
Class 1, Type B, Ip21 |
||
Working Environment |
Temperature : 10 to 40°c, Relative Humidity : 10 to 90% Non Condensing |
||
Configuration |
Basic: Adult-Pediatric, Neo: Neonatal - Pediatric, Universal: Adult-Neonatal, Advance*: Adult-Pead-Neonatal |
Regarding any queries.