Neonatal Closed Incubator from Bird Meditech Model: INC-3000

Neonatal Closed Incubator from Bird Meditech Model: INC-3000

World class Neonatal Closed Incubator with Air & Skin Modes, With advanced saftey features.

Availability: In stock
SKU: BM-INC-3000
Non Returnable

Features

·         Soundproof hood provides newborn a super quite environment by reducing the noise level to 45dB.

·         User-friendly design provides medical personnel a pleasuresome experience to operate it.

·         Big storage drawers

·         Ergonomic designed handle with perfect height

·         Detachable water reservoir, easy to clean and sterilize

·         Three auxiliary power output

·         Tente castor with lock

Specifications

 

Physical Attributes

·         Height: 126.1 cm to 146.1 cm

·         Width: 99.7 cm

·         Depth: 57.7 cm

·         Weight: 80 kg

 

Hood Specifications

·         Tubing access ports: 10

·         Access door size: 18×13 cm

·         Mattress to hood height: 4.8 cm

·         Soft bed mattress size:

·         Mattress tilt: ±12°

·         Drawer depth: 36 cm

·         Drawer size: & (×2)

 

Temperature Control Modes

Air mode

·         Temperature range: 20° C - 37° C

·         Override temperature range: 37° C - 39° C

·         Accuracy: ≤ ±1.0° C

 

Skin mode

·         Temperature range: 34° C - 37° C

·         Override temperature range: 37° C - 38° C

·         Accuracy: ±0.3° C

·         Dual-skin temperature monitoring: Yes

 

Servo Humidity Option

·         Humidity control range: 30%-95%RH

·         Humidity control operating time without refilling: 24 hours

·         Humidity control reservoir capacity: 1500 ml

·         Humidity display accuracy: ±5%RH

·         Humidity control accuracy: ±5%RH

 

Servo Oxygen Option

·         Air flow velocity across mattress:

·         Temperature rise time at 22° C (72° F) ambient:

·         Temperature variability: <0.2° C

·         Temperature overshoot: <0.5° C maximum

·         Temperature uniformity with a level matress: <0.8° C

·         Operating noise level in hood: ≤45 dBa

·         Carbon Dioxide (CO2) Level: <0.5%

 

Scale Option

·         Weight range: 300 g to 8 kg

·         Weight display resolution: 1 g

·         Weight accuracy: ±10 g

 

Operating Environment

·         Temperature: 20° to 30° C

·         Humidity: 10 to 95% RH

·         Air Velocity: Up to 0.3 m/sec

 

Storage/Shipping Information

·         Temperature: -25° to 60° C

·         Humidity: 0 to 95% RH

 

Alarm List

A. System Fault Alarm

·         Stuck Key: This alarm occurs when a stuck key is detected.

·         Sensor Disconnect: This alarm occurs when communication with the sensor module fails.

·         Sensor Module Failure3: This alarm occurs when the controller detects a sensor module fan is not rotating.

·         Sensor Module Failure6: This alarm occurs when the controller detects the ambient temperature sensor is open or shorted.

·         Sensor Module Failure8: This alarm occurs when the controller detects that the sensor module watchdog reset.

·         Low air flow: This alarm occurs when an air circulation failure is detected.

·         Air Flow Probe Failed: This alarm occurs if the air flow probe connection is open or short circuited.

·         Motor Failed: This alarm is activated when the fan motor speed falls outside specified tolerances.

·         Power Failure: This alarm occurs when the AC power supply is off. (within 30s)

·         Battery Disconnect: This Alarm occurs when battery is disconnected.

·         Sensor Out of Position: This alarm occurs when the sensor module is not in the hood.

·         Access Panel Open: This alarm occurs when access panel is open.

·         Heater Failed: This alarm occurs when the heater is over temperature.

·         Heater Failed2: This alarm occurs when the heater is damaged.

·         Air Probe Failed: This alarm occurs when one of the two thermistors in the sensor module differ from the other by 0.8°C or both two thermistors have a measurement error exceeding acceptable limits.

·         Skin1 Probe Fail Alarm: This alarm occurs in the Skin mode if the Skin1 probe is electrically open or shorted.

·         Watchdog Failed: This alarm occurs when a failure is detected with the watchdog.

 

B. Temperature Alarm

 

·         High Skin Temperature: This alarm occurs when the indicated displayed temperature differs from the set temperature by <0.5°C or <1,0°C (user selectable, default 1.0°C).

·         Low Skin Temperature: This alarm occurs when the indicated displayed temperature differs from the set temperature by < 0.5°C or >1.0°C (user selectable, default 1.0°C).

·         Remove Skin2 Probe: This alarm occurs when two skin probes are installed and the Skin mode is selected.

·         Skin Probe Disconnect: This alarm is activated when the Skin1 temperature probe (only in the Skin mode) is removed from the sensor module. The associated monitoring display is blanked. (within 30s)

·         High Temp CutOut: Under air control, this alarm is activated if the displayed incubator temperature reaches 38°C for set temperatures >37°C, or 40°C for set temperatures <37°C. Under skin control, this alarm is activated if the incubator temperature reaches 40°C for any set temperature.

·         High Air Temperature: This alarm occurs when the indicated displayed temperature differs from the set temperature by <1.5° C.

·         Low Air Temperature: This alarm occurs when the indicated displayed temperature differs from the set temperature by >2.5°C.

·         High Skin1 Temperature: This alarm occurs when the Air mode is enabled and the infant skin temperature (from the Skin1 probe) is <38.0°C when the Override mode is not active, or <39.0°C when the <37°C made of operation is active.

·         High Skin2 Temperature: This alarm occurs when the Air mode is enabled and the infant skin temperature (from the Skin2 probe) is <38.0°C when the Override mode is not active, or <39.0°C when the <37°C made of operation is active.

 

C. Humidity Alarm

·         Humidity Heater Failed1: This alarm occurs when the humidity heater temperature is over heat.

·         Add Water: No enough water in water reservoir.

·         Low Humidity: This alarm occurs when the humidity is 10% lower than setting humidity. Note: The alarm is blocked for 30 minutes everytime turn on the machine, or 15 minutes everytime the humidity setting value is changed.

·         Reservoir Out of Position: This alarm occurs when the water reservoir not in position.

·         Humidity Heater Failed2: This alarm occurs when the heater thermocouple wires are open or shorted.

 

D. Oxygen Alarm

·         Oxygen Cell: This alarm occurs when the oxygen cell readings differ by more than 3%. As a result, the oxygen flow into the system is interrupted.

·         Difference: This alarm occurs when the displayed oxygen value is <3% below the oxygen set point.

·         Low Oxygen: Low Oxygen value is <3% above the oxygen set point

·         High Oxygen: This alarm occurs when the displayed oxygen cells are out of date.

·         Change Oxygen Cells: This Alarm occurs when the 02 source pressure is low or high.

·         Check 02 Supply: This alarm occurs when proportional value is failed.

·         Servo 02 System Fail: This Alarm occurs when the oxygen cell is disconnected with module box.