Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
Bi Phasic Defibrillator with AED, Adult & Pediatric paddles , ECG Monitor, Recorder
Standard Configuration
AED, RESP, Thermal Recorder, 360ΩJ Optional Modes – Pacer, NIBP, EtCO2, SpO2
Physical Specification
Size : 295mm x 252mm x 316mm
Weight : 6.5 kg.
Screen Size : 7 Inch TFT Screen
Resolution : 800x480
Waveforms : Max 4 waveforms
Environmental Configuration
Temp : 0–45°C
Humidity : 10%–95%, non-condensation Ingress Protection IP44
Power Requirement : 100–240V, 50/60Hz±3Hz
Battery
Battery : Lithium-ion Battery
Battery Backup : Monitoring Mode - min. 6hrs
Defib Mode : 210 Shocks Pacing Mode: 4.5 hours
Data Storage
Alarm Event : 200groups
Wave Review : 10mins
NIBP Review : 2000groups
Trend Graph : 160hrs
Trend Table : 160hrs
Voice Recording : Max 240mins in total
Defibrillator
Operating Mode : Manual Mode, AED Mode, Synchronous Defibrillation
Waveform : Bi-phasic, Truncated exponential waveform with impedance compensation
Battery Power : Less than 7 seconds to 360joules with a new, fully charged battery
Charge Time : Less than 8 seconds to 360 Joules (AC Power)
Patience Impedance : 20–300Ω (External defibrillation)
Range
Manual Mode : 1J–360J, 25 types
(1/2/3/4/5/6/7/8/9/10/15/20/30/50/70/100/150/170/ 200/220/250/270/300/360J)
Noninvasive Pacing (Optional)
Waveform : Moniphasic square wave pulse
Pulse Width : 20ms or 40ms
Accuracy : ±15%
Pacing Mode : On-demand or fixed
Pacing Frequency : 30ppm to 210ppm
Accuracy : ±1ppm or ±1.5% (whichever is greater)
Pacing Output : 0 mA to 200 mA
Accuracy : ±5% or ±5mA, whichever is greater
Monitoring
Lead Type : 3 leads ECG, 5 leads ECG, AUTO
Sweep Speed 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
Heart Rate : Adult : 15–300bpm; Paediatric: 15–350bpm
Resolution : 1bpm
Accuracy : ±1bpm
CMRR
Monitoring : >105dB
Diagnosis: > 90dB
Surgery: >105dB ST: >105dB
Recorder
Type : Built-in; Thermal Array
Channel : Max 3 channel waveforms
Respiration
Method : Thoracic Impedance Method
RR Measurement : Adult: 0–120rpm; Pediatric: 0–150bpm
Accuracy : 7–150rpm: ±2rpm or ±2% (whichever is greater) 0–6rpm: unspecified
Interfacing
USB Interface RJ45 interface
Multi-functional connector
Optionals
SpO2
Type : Self / Nellcor / Masimo
Range : 0–100%
Resolution : 1%
Accuracy : ±2%
PR Range : 20–254bpm
NIBP
Initial Inflation Range : 20–250 mmHG
Over Pressure Protection : 300mmHG
Resolution : ±3mmHG
Range of Systolic Pressure : 40–260mmHG (Adult) / 40–160mmHG (Paed)
Range of Dystolic Pressure : 20–200mmHG (Adult) / 20–120mmHG (Paed)
Range of Mean Pressure : 26–220mmHG (Adult) / 26–133mmHG (Paed)
PR from NIBP : 30–220bpm
AED Mode
Output Energy : Adjustable; 100J–360J
Number of Electric Shock : Adjustable; Once / Twice / 3 Times
Types of AED : VF & VT
EtCO2
Type : Respronics / Masimo
Measurement Range : 0–190mmHG
Resolution : 1mmHG or 0.1% or 0.1kPa
AWRR Range : 0–150rpm
AWRR Accuracy : ±1rpm