Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
CUT MODE
PURE - A high-frequency
alternating current yields smooth, rapid cuts that evoke little to no
hemostasis. BLEND 1 - Preset blended cut modes give the surgeon varying degrees
of hemostasis in cut mode.
BLEND 2 - The mode is used for
the vaporization of prostatic tissue or any fat tissue. The higher output
wattage gives fast cutting e_ect of the tissue.
ENDOCUT - The mode is used to
automatically control cut system to reduce the complication rate of endoscopic sphincterotomy
(EST) and serum hyperamylasemia after EST compared to the conventional blended
cut mode.
COAG MODE
SPRAY - Spray output is made
for direct coagulation at the tissue with spark. This spark does not give any
cutting e_ect as its crest factor is high.
FORCE - This mode is a greater
degree of homeostasis and covers all demands of standard coagulation. This
produce the fine tune waveform that creates superficial resistance in the tissue
resulting in superior coagulation where the tissue destroyed by hyper
evaporation.
FULGURATE
- Sparking
the tissue to lead to coagulation. There is no tissue-electrode contact;
rather, voltage is raised in order to incite a spark between electrodes in
order to coagulate the tissue in between.
DESICCATE
- Coagulation
in which the active electrode is in direct contact with the tissue is referred
to as desiccation and is the type of coagulation used in most surgeries It
ensures pinpoint desiccation with less destruction of peripheral tissue.
BIPOLAR
MODE
MICRO - High Voltage force
coagulation for special purpose when speed is required.
MACRO - For highly precise
control over coagulation process and privations of carbonization.
CUT - This Bipolar CUT
technology nearly eliminate excessive current, even during precision cutting in
minute tissue structure.
AUTO - Coagulation start by
pressing the footswitch and stops output and audio signal automatically when
the tissue is coagulated.
Regarding any queries.