Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
Specifications:
· Power Supply (I/P) : AC 220V to 240V 50Hz
· Battery : 12V NiMH Battery
· Battery Backup : 10hrs @ 5ml.hr Flow rate on 50ml Syringe
· Power Consumption : 7 VA
· Applicable Syringe : 5ml, 10ml, 20ml, 30ml, 50ml &100ml Syringes from JMS, Braun, Dispovan, BD and
· Custom Programmable for any Brand of Syringe.
· Automatic Syringe Detector : Available
· Custom Programmable Syringe : Available
· Flow Rate Range :
a) 0.1 – 200 ml/Hr 70ml/Hr (5ml Syringe)
b) 0.1 – 300 ml/Hr 110ml/Hr (10ml Syringe)
c) 0.1 – 500 ml/Hr 190ml/Hr (20ml Syringe)
d) 0.1 – 700 ml/Hr 260ml/Hr (30ml Syringe)
e) 0.1 – 1200 ml/Hr 400ml/Hr (50ml Syringe)
· Infused Volume : 0.1 – 999.9 ml
· Elapsed Time : Display of Elapsed time from 00:01 to 99:59 Hrs
· Modes : Drug Library mode, Flow Rate only, Flow Rate & Target volume, Flow Rate & Target Time Target volume and target time
· Bolus : Purge Bolus & Auto Bolus
· Bolus Flow Rate :
· 10 – 200 ml/Hr 70 ml/Hr (5ml Syringe)
· 10 – 300 ml/Hr 110 ml/Hr (10ml Syringe)
· 10 – 500 ml/Hr 190 ml/Hr (20ml Syringe)
· 10 – 700 ml/Hr 260 ml/Hr (30ml Syringe)
· 10 – 1200 ml/Hr 400 ml/Hr (50ml Syringe)
· Bolus Volume : 0.1ml to 10 ml in 0.1 ml increments
· Dosage Taper Up/ Down(Sequential Mode) : Available with 10 different programs
· KVO : 0.1ml/Hr to 9.9ml/Hr in 0.1 ml increment
· Occlusion Pressure : 10 steps selection (OFF to 9 Levels)
· Alarms : tAr.1 – Target Volume reached
tAr.2 – Target time reached
End – End of Syringe
occL – Occlusion
OFF – Battery discharged
· AL 1 – Syringe holder disturbed during infusion
· AL 2 – Syringe Disengaged while running
· AL 3 – Motor unable to drive
· AL 4 – Mains to Battery F/R High reduce F/R 200
· AL 5 – AC Power over limit
· AL 6 – AC Voltage below 215
· AL 9 – Syringe not engage
· Audio/Visual Indication : Target reached (time or volume), End of syringe, Occlusion, Battery discharged, Motor unable to drive, Syringe not engaged, Near Empty, Battery low, Operation on mains / battery
· Display : Four 7 segment Display
· PCA (Optional) : PCA dose can be preset from 0.1 to 10ml. Minimum locking time interval between two doses,
· Time interval from 0 -24Hrs in steps of 1 min, Delivered and non delivered doses display
· Accuracy : Mechanical Accuracy ±1%; Including Syringe ± 3%
· Safety Testing : EN 60601-1, EN 60601-1-2, EN 60601-1-6, EN 60601-1-8, EN 60601-2-24
· Certification : ISO 9001:2008; ISO 13485:2003; CE
· Dimension : 170 x 107 x 135 ( LxBxW)
· Space Occupied : 2.4 ltrs
· Weight : 1.75kg
Regarding any queries.