Vessel Sealing Mode: Sealing up to 7mm & energy stops with audio indication at completion of sealing cycle.
Saline Resection Mode: Saline Plasma Bipolar Cut & Coag with reusable Bipolar working element & loops.
Bipolar Cut & Coag Mode: Bipolar Coaglution & Cutting of Tissue by Shalya Bipolar Scalpel (Desiccator)
Dual Mono (Optional) : Two Monopolar output channel with independent control over each output for ON & OFF both channels work by foot as well as hand control.
6SENSE Technology : Maintain set power over wider range of tissue by Tissue Feedback Technology.
PREM : Patient Return Electrode Monitoring gives an Alarm providing safety against the Return Electrode site burns.
Tissue ReHold Alarm : “CL rEHOLd” (Close Rehold) is given if Insert is closed & “OP rEHOLd” (Open Rehold) in case of open.
Randomized SPRAY Coag : Homogeneous, superficial, faster Coagulation over larger area.
Self Diagnosis : Ensure the ESU with accessories are safe for functioning & display Error Codes if any.
Upgradability : Argon Beam Coagulation, Ultrasonic Aspirator, Smoke Evacuator, Suction Irrigation module.
User Programmable : THIRTY Surgical Programs help in setting the system after power ON.
BiCoag Alarm : Audio feedback after completion of Bipolar Coag, reduces charring and sticking.
Pulse/Endo Cut : Tissue feedback Pulse Interval controlled Pulse Cut for Polypectomy & Papillotomy.
Remote Power Settings (Optional) : Change mode and power from sterile field by normal Handswitch Pencil.
AutoBip (Optional) : Automatic Start & Stop of Bipolar current by tissue sensing without foot switch activation.
Foot Switches / Pedals : Explosion Proof (Fire Resistance) and Water Proof Foot Pedals
Time Out Function (Optional) : Automatically stops the HF output if prolong unintended activation.
Surgical Applications : Useful for wide range of surgical application as : Plastic, Neuro, ENT, Spine, Opthal, Laparo Surgery, Endo, GI, Ortho, Open & General Surgery, Onco, Gynecology, Hernia, GI Endocut, Papillotomy, Polypectomy, Under Water, Urology, Gynec TCRE, Hysteroscopy, Thoracotomy, Cardio Thoracic.
Vessel Sealing: Laparoscopic / Radical / Vaginal/ Abdominal Hysterectomy, Appendectomy, Colectomy, Cystectomy, Gastric , Bypass, Hemorhoidectomy, LAVH, Liver Resection, Nephrectomy, Nissen Fundoplication, Splenectomy.
Saline Plasma Bipolar Resection: Bipolar TURP/ TCRE / Hysteroscopy / Arthroscopic Surgery
MONOPOLAR CUT: Pure Sine wave 390 kHz
Sp Cut - Continues output, 300 W at 300 Ω, CF 1.5
Pure - Continues output, 300 W at 300 Ω, CF 1.5
Blend - Bursts repeating at 29 kHz 55% ON cycle, 200 W at 300 Ω, CF 2.5
Endo - Pulse output, 300W at 300 Ω, CF 1.5
BIPOLAR: Pure Sine wave 440 kHz
Micro -Lower output voltage, 95 W at 100 Ω, CF 1.5
Macro - Higher output voltage, 95 W at 100 Ω, CF 1.5
SALINE RESECTION: SIX Modes
Saline Cut Modes: 390 kHz
TURoS Cut : Continues output 300 W at 100 Ω, CF 1.5, High Volt
Saline Coag Modes: 440kHz
TURoS Coag: Continues output 200 W at 100 Ω, CF 1.5
FEEDBACK SYSTEM
• 3 Microprocessor Controlled 6SENSE Technology with sampling rate above 4,00,000 sense/second
• Maximum output power deviation < 15% or 5 watts, whichever is greater at any setting with PER-98
LEAKAGE CURRENTS
• RF leakage current:- Monopolar : < 150 mA rms; Bipolar : < 60 mA rms
• Low Frequency leakage currents
• Normal polarity, intact ground : < 10 μA
• Normal polarity, ground open : < 50 μA
MONOPOLAR COAG : Damped Sine wave 460 kHz
Soft -Repetition Frequency 56 kHz, 120 W at 500 Ω, CF 5.0
Fulgurate LCF -Repetition Frequency 36 kHz, 120 W at 500 Ω, CF 6.0
Fulgurate HCF-Repetition Frequency 29 kHz, 120 W at 500 Ω, CF 7.0
Randomized Spray - Randomized Repetition Frequency 25 kHz < f < 35 kHz, 120 W at 500 Ω, CF 8.3
VESSEL SEALING: Pure Sine wave 440 kHz
Seal - Pulse output mode, 5 level, 150 W at 100 Ω, CF 1.5,
Bip Coag (Standard) - Medium output voltage, 95 W 100 Ω, CF 1.5,
Bip Cut (Macro) - Higher output voltage, 95 W 100 Ω, CF 1.5
PREM SYSTEM
Measuring Frequency : 80 kHz ± 10 kHz
Measuring current : < 10 μA Acceptable Resistance Rang
Dual Area PREM Patient Return Electrode : 5-135 Ω (Adaptive)
Single Area Patient Return Electrode : < 10 Ω
DESIGN STANDARD
Classification according to the EC- Directive 93/ 42/ EEC: Class IIb
General Safety Standards: IEC 60601-1-1, IEC 60601-2-2
EMI/EMC Compatibility Standard: IEC 60601-1-2
IS Safety Standard: IS 7583
INPUT POWER (other ratings available on request)
Operating Voltage : 180-265 Volt AC, 50-60 Hz
Max. Power input : 1000 VA, Fuse : 6 A, UPS Backup : 1.5KVA
DIMENSIONS: L x W x H in cm - 45.0 x 35.5 x 11.5, Weight in kg < 7.5