· Anti Hypoxic pneumatic device. Ensure no nitrous oxide flow, if oxygen flow is lower than 0.5 Ltr / min approx.
· Minimum 25% (+/-5%) oxygen in surgery gas mixture
· Ration control 1:3 mixed ratio between O2, & N2O gas flow.
· Rota meter 9"
· Oxygen driven N2O supply.
· Audio visual oxy failure warning device (OFWD)
· Higher efficiency forged regulators & yokes with SS fittings.
· High precision pressure relief cum non return valve
· Vaporizer Gold Man Halothane
· Change over mechanism with oxygen flush. (open & close circuit)
· Gas inlet Port O2 & N2O each 3 Nos
· Pin-indexed gas cylinder connections for O2 & N2O each 2 Nos & diameter index gas specific central pipe line connection each one no. for O2 & N2O
· Standard Mogills circuit all rubber antistatic tested.
· Trolley mode of M.S multi colour power coated.
· Stainless steel table top tray
· Modified monitoring accessories tray, drawer top stainless steel tray
· 5" dia antistatic breaking castor wheel
· Oxygen outlet for driving ventilator (double)
· Large dia (63mm) panel mounted cylinder pressure, operating pressure & pipeline pressure display gauges for O2 & N2O
· 10 Ltr. W/C cylinder carrier facility. Provision for tec vaporizer, circle absorber & ventilator. (optional at extra cost)
· Size 1530 mm x 900 mm x 770 mm (H x W x D)