Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
Quality
Standards and Classification
·
Anti-electroshock degree: Class-Ⅰ
equipment (internal power supply)
·
TEMP/SpO 2 /NIBP: BF
·
ECG/Resp: CF
Application Range
·
Adult/ Pediatric/ Neonatal/ Medicine/
Surgery/ Operating Room/ ICU/ CCU
Display
·
12.1” real color TFT screen
Environment
working
environment conditions
·
Temperature: Working
temperature 0 - 40 ℃
·
Humidity: Working humidity ≤
85 %
Shipping
and storage conditions
·
Temperature: Transportation
and storage temperature -20 - 60 ℃
·
Humidity: Transportation and
storage humidity ≤ 93 %
Electrical Specifications
·
Power supply voltage: AC
100~240V
·
AC, power frequency: 50/60Hz, maximum
input power 70VA Fuse is F2AL 250V
·
Working method: continuous
work
·
Built-in lithium battery:
d.c11.1V not less than 2000mA, working time not less than 2 hours.
Dimension and Weight
·
Product Size 262*336*84mm
·
product weight 2.72kg
ECG
·
Gain: X0.25, X0.5, X1, X2,
Auto
·
Waveform speed: 12.5mm/s,
25mm/s, 50mm/s,
·
Sensitivity>200uV
(peak-to-peak)
Bandwidth:
·
Diagnostic mode: 0.05~130Hz
·
Monitoring mode: 0.5~40Hz
·
Operation mode: 1~20Hz
·
Common Mode Rejection Ratio:
·
Diagnostic mode: >90dB
·
Monitoring mode: >100dB
·
Operation mode: >100dB
·
Electrode polarization
voltage: ±300mV
·
Input impedance≥5M
·
DC bias voltage: ±300mV
·
Calibration signal: 1mV
(peak-to-peak), accuracy ±5%
ST
segment detection measurement:
·
Measurement range:
-2.0mV~+2.0mV
·
Measurement accuracy: within
the range of -0.8mV~+0.8mV, the measurement error is ±0.02mV or ±10%, whichever
is greater, other ranges are undefined
Trend data length
·
School limit list:400
groups
·
Waveform playback:6000
seconds
·
Trend table:7
days
·
Trend graph:up
to 7 days
·
Alarm review:60
groups
ST Segment
·
Maximum upper limit 2.0mv
·
Minimum lower limit -2.0mv
·
Single adjustment amount 0.1mv
NIBP
·
Measuring Range and Precision Range
·
Adult Systolic blood pressure
40~270mmHg
·
Diastolic blood pressure 10~215mmHg
·
Mean blood pressure 20~235mmHg
·
Static pressure range 0~300mmHg
·
Static pressure precision ±3mmHg
Pressure
precision:
·
Max. average error: ±5mmHg;
·
Max. standard deviation: 8mmHg
SpO2
·
Measuring range 0~100%
·
Resolution 1%
·
The accuracy is within the
range of 70% to 100%, and the measurement error should be ±2%; In the range of
0% to 69%, the measurement error is Undefined
Pulse Rate
·
Measuring range 30~300bpm
·
Resolution 1bpm
·
Accuracy ±3bpm
·