Working Hours: 10AM - 6 PM (or)
Fill Service Request Form
Structure: Square tubular Electrostatic powder coated steel sections.
Top Tray: Rigid top tray for monitor
Table Top: Stainless Steel
Drawer: One/ Two large storage drawer
Cylinder Yokes: Gas specific yokes & gauges with clamping bars, dual for Oxygen and one for Nitrous Oxide.
Mobility: Four large diameters anti static caster wheels with ball bearings, front castors have brakes.
Pressure Gauges: Two for O2 & One for N2O cylinders pressure & one O2 line pressure.
Regulators: For higher efficiency & reliability two for Oxygen & One for Nitrous oxide.
Non-return cum pressure release valve: Minimizes risk of back flow of gases, blows off when pressure exceeds 200 cm H2O.
Oxygen Failure: Pneumatic device which activates audible alarm (for minimum 10 seconds) when Oxygen supply pressure falls to 205 kPa.
Vaporizers: Space for One Vaporizer provided as standard. Vaporizers are optional.
Flow meters: Long (upto 260mm) rotating bobbin flow meters calibrated in double/ triple scale, antistatic & non-interchangeable:-
Oxygen being a preferred gas kept on downstream of other Gas.
Oxygen Flush: Emergency Oxygen flush outlet button at table level (36-50 lit/min.)
Change Over Unit: Change over unit is provided for open or close circuit.
Breathing Circuit: Standard Bain Circuit.
Nitrous Cut-off: Provide Nitrous Cut-off system prevents risk of administrating 100% Nitrous Oxide in absence of Oxygen pressure.
Standard Accessories: One / Two built-in Oxygen outlets (422 kPa) for driving ventilators, key spanner for A type cylinders, face mask for Adult & Pediatric.
Optional (Available at extra cost):
Regarding any queries.